The following has been reported:
biomed reported: continuous beeping and flashing a red light.Screen was showing an error message, but it was not recorded.Touch screen was unresponsive.Unit was forced to shut down by holding down the power button for 1 minute.Turned on after and behaved normally, however still sending to vendor to investigate previous unusual behavior.A preliminary review identified the following issue: processor hardware failure (linux core)
an active infusion was stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
|