| Model Number |
1054260B |
| Medical Device Problem Codes |
Failure to Calibrate (2440); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The trilogy 100 was returned to the manufacturer service center for service.There was no serious patient harm or injury that occurred or was reported.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the service center, the device failed a vac power source test step.The power management board was replaced to address the test step failure.The device passed all testing during the retesting.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of vac power source is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device exceeds its expected life.
|
| |
|