| Catalog Number |
08057931190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with iron gen.2 on a cobas pure c 303 analytical unit.No questionable results were reported outside of the laboratory.The sample initially resulted in an iron value of 23.9 mg/dl.Per laboratory protocol, the customer repeats all samples that initially result in an iron value of < 25 mg/dl.The sample was repeated and the result was 23.9 mg/dl.Since the results did not match the patient's history, the sample was repeated two additional times.The additional iron repeat values were 36.3 mg/dl and 39 mg/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
The serial number of the c 303 analyzer is (b)(6).The field service engineer cleaned the water tank, probes, and syringe lines.Calibration and controls passed.Patient samples and precision studies were run and no further discrepancies were observed.The investigation is ongoing.
|
| |
|