| Catalog Number |
0042-0000-US |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The complainant reported that during service and repair activities, an automated impella controller (aic) underwent preventive maintenance and functional testing.During the purge cassette portion of the preventive maintenance procedure, the console failed testing when a crack was identified in the purge disk retainer.The purge disk retainer was replaced.Following replacement, the preventive maintenance procedure was completed without further issues.The console subsequently passed testing and was returned to service.The issue was identified during service and repair activities and not during clinical use.No patient involvement was reported in association with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|