| Catalog Number |
1000323 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pump was originally pre-closed with per close two times.Upon insertion abiomed representative instructed the physician that per close was not part of our ifu recommendations.The physician insisted on placing two anyway with acknowledgment of the risks.Upon successful wean and attempt to explant after removal of the repositioning sheath, there was resistance met at the motor housing portion of the pump removal.It was determined by the physician that the per close devices had been tightened around the catheter, and therefore the vein puncture site was not large enough to fit the motor out of the vein.The abiomed representative was notified and they discussed that the safest option would more than likely be a cut down to remove the pump safely from the patient.Ultimately the physician was able to break the per close knot and remove the pump safely from the patient and removed the pump safely from the patient.No allegations were made against the device or the devices function with the doctor acknowledging that it was the result of the physician tightening the knots.Patient safe and stable at end of procedure.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|