| Model Number |
CF-HQ190L |
| Medical Device Problem Codes |
Contamination (1120); No Display/Image (1183); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the colonovideoscope displayed no video output during a diagnostic colonoscopy while the patient was sedated and the device was in use.Upon connection, the image would briefly appear before the image shut off.The procedure was completed using the same device.During the initial evaluation of the device, it was observed there was an error code e216.In addition, it was also noted there was white residue inside the air / water (a/w) channel and inside the auxiliary water flow.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The device was returned to olympus for inspection, and the reported failure along with the initial evaluation findings were confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely that the reported no-image condition, along with the initial evaluation finding of error code e216, was caused by an expected or random component failure without any design or manufacturing issue.The cause of the initial evaluation findings of white residue in the air/water channel and auxiliary channel could not be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device.
|
| |
|