| Model Number |
1054260B |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy 100 ventilator device was returned to a third-party service center for service.There is no allegation of serious or permanent harm or injury.The device was not in use on a patient at the time of the occurrence.During the evaluation, the service technician observed r ventilator inoperative error code e009.(err_dsp_motor_failure) in the error log.No documentation of a replaced component to address the issue.Additionally, no visible foam particles were identified.The device was scrapped; therefore, no additional repair information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Dhr review was performed for the complaint device serial number, (b)(6) review confirms that the device met the final acceptance criteria and was released for final distribution.The reported failure of r ventilator inoperative error is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|
Search Alerts/Recalls
|