• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Respironics, Inc. Trilogy 100; Ventilator, Continuous, Facility Use

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Respironics, Inc. Trilogy 100; Ventilator, Continuous, Facility Use Back to Search Results
Model Number 1054260B
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2026
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy 100 ventilator device was returned to a third-party service center for service.There is no allegation of serious or permanent harm or injury.The device was not in use on a patient at the time of the occurrence.During the evaluation, the service technician observed r ventilator inoperative error code e009.(err_dsp_motor_failure) in the error log.No documentation of a replaced component to address the issue.Additionally, no visible foam particles were identified.The device was scrapped; therefore, no additional repair information was provided.
 
Additional Manufacturer Narrative
Dhr review was performed for the complaint device serial number, (b)(6) review confirms that the device met the final acceptance criteria and was released for final distribution.The reported failure of r ventilator inoperative error is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Trilogy 100
Common Device Name
Ventilator, Continuous, Facility Use
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50948176
Report Number2518422-2026-046655
Device Sequence Number16063975
Product Code CBK
UDI-Device Identifier00606959026353
UDI-Public(01)00606959026353
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1054260B
Device Catalogue Number1054260B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/31/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-