| Model Number |
380747-41 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
There were two procedures identified for the event date.Review of the system logs found no errors during either procedure.Review of the system logs from five subsequent procedures performed on the system also found no errors.An intuitive surgical medical safety officer (mso) reviewed the event and concluded that the patient in this report died following a robotic procedure in which the patient had a post-operative complication of an incarcerated hernia from a port site.While the port sites are used for the robotic procedure, these sites need to be closed after surgery with a surgical process that does not include the use of the robot.Based on the information provided in the summary of events, no intuitive surgical product or instrument contributed to this event.
|
| |
|
Event or Problem Description
|
|
It was reported that following a da vinci-assisted hiatal hernia procedure, the patient developed an incarcerated trocar-site hernia and subsequently expired.On postoperative day five, the patient presented to the emergency room, where the hernia was identified and an unspecified subsequent procedure was performed.The cause and date of death, as well as additional details regarding the events leading to the death, were not provided.Attempts to obtain additional information from the surgeon and site were unsuccessful.
|
| |
|