Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735943, serial/lot #: (b)(6), ubd: , udi#: h3, h6: a manufacturer representative went to the site to test the equipment.In summary, the power cable was starting to show some signs of wear on the outer coating of the cable.The power cable was replaced.Codes fdm b01, fdr c0201, fdc d02 are applicable to this analysis.D9, h3, h6: the hardware was returned and analysis was performed.The returned power cable did have a damaged cable jacket, but it didn't have exposed wires.The cable passed continuity testing.Codes fdm b01, fdr c0201, fdc d02 are applicable to this analysis.H6: multiple fdd/annex a codes were reported.A040507 was coded for the component being worn.A07 was coded for the grey coating changing to white.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|