| Model Number |
97715 |
| Medical Device Problem Codes |
High impedance (1291); Unintended Collision (1429)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from a manufacturer's representative (rep) regarding an implantable neurostimulator (ins).The rep stated the patient (pt) was explanted this morning due to high impedances and also due to the fact that the pt had a fall, all starting in (b)(6) 2026.The rep stated the healthcare provider (hcp) specified that there was potential for fracture of the leads.The leads and battery were removed and 2 new leads and a battery were placed. **see pe#709415900 for split information regarding new implant.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date is approximate.Month and year are confirmed valid.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|