| Model Number |
A219 |
| Medical Device Problem Codes |
Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
04/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had inappropriate shocks due to t-wave oversensing via reduced intrinsic complexes.The smart pass feature had programmed itself off due to the low intrinsic amplitudes.The sensing vector was programmed to the primary vector.Office testing found the sensing was poor in all three of the sensing vectors.The xrays were done and are unchanged from implant.The electrode is a bit left of the sternum to avoid the sternal wires.Technical services advised some testing options.There were no additional adverse patient effects.The system remains implanted.
|
| |
|