| Model Number |
G56572 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an unknown endoscopic procedure, the physician used a cook hemospray endoscopic hemostat.It was reported that they had a problem with the release of hemospray powder.They managed to make a first release of dust "[spray]" but then had no more co2 "[carbon dioxide]".At the time of this report, our attempts to collect additional information regarding patient outcome have been unsuccessful.While the complainant did not specify if the patient experienced any adverse effects or required additional medical procedures due to this event, the information able to be collected does not reasonably suggest the patient was adversely impacted.Additional attempts to gather this information will be made.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued:
section e: phone: (b)(6).Investigation evaluation:
a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A photo was provided and reviewed.The lot number in the photo matches this report.The label in the photo matches the product reported.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion:
we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.A corrective action (capa) was initiated to further investigate device failure due to being unable to spray powder.This device is within the scope of the capa.Prior to distribution, all hemospray endoscopic hemostats are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action:
a corrective action (capa) was initiated to further investigate device failure due to being unable to spray powder.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Search Alerts/Recalls
|