Olympus Winter & Ibe GmbH Video telescope "ENDOEYE HD II", 10 mm, 30°, autoclavable; Laparoscope, Gynecologic
|
Back to Search Results |
|
| Model Number |
WA50042A |
| Medical Device Problem Codes |
Poor Quality Image (1408); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the image appears green involving an olympus laparoscope, gynecologic.Additionally, during the initial evaluation of the device, damaged fibre bonding in the outer tube area was observed.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection, and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation regarding the allegation, the cause was a broken video cable.For damaged fiber bonding in the outer tube area, the cause was not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|