• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; MONITOR,BP,WRIST,DIGITAL UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; MONITOR,BP,WRIST,DIGITAL UNIT Back to Search Results
Catalog Number MDS4003
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "unit is displaying inaccurate numbers.Checked versus larger non portable unit.Looks like calibration is wrong.".
 
Additional Manufacturer Narrative
The unit was displaying inaccurate numbers, checked versus a larger non portable unit, and it looked like calibration was wrong.Medline received 1 automatic digital blood pressure monitor; the device was tested under simulated conditions of 160/100 @ 80 bpm and 120/80 @ 80bpm, and the device had readings of 154/100 @ 80 bpm and 121/80 @ 80 bpm with no error message observed in testing.The complaint was not confirmed, with cause code not confirmed with received sample.No further action was taken because objectionable trends were not seen for this issue, and the division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
MONITOR,BP,WRIST,DIGITAL UNIT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
herchelle teodoro
three lakes drive
northfield, IL 60093--275
MDR Report Key50948597
Report Number1417592-2026-54623
Device Sequence Number10452072
Product Code DXN
UDI-Device Identifier00888277675493
UDI-Public00888277675493
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K122482
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS4003
Device Lot Number07521120002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2023
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-