The unit was displaying inaccurate numbers, checked versus a larger non portable unit, and it looked like calibration was wrong.Medline received 1 automatic digital blood pressure monitor; the device was tested under simulated conditions of 160/100 @ 80 bpm and 120/80 @ 80bpm, and the device had readings of 154/100 @ 80 bpm and 121/80 @ 80 bpm with no error message observed in testing.The complaint was not confirmed, with cause code not confirmed with received sample.No further action was taken because objectionable trends were not seen for this issue, and the division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|