It was reported that during in-house engineering evaluation, it was determined that the fms vue ii pump-shaver box device had a leak ¿ liquid.It was initially reported that preoperatively to an unknown procedure, it was observed that the device could not be turned on because there was moisture inside the device.It was reported that there were no delays to the surgical procedure as a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative:
investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified:
functional: moisture/water damage, visual: cracked, functional: does not function, functional: leak - liquid and functional: electrical/electronic component failure.Per service report, this complaint was confirmed.The control valve, switch/relay, fuse and circuit board were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a manufacturing record evaluation was performed for the finished device serial number ((b)(6)), and no non-conformance was identified.
|