• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; MONITOR,BP,WRIST,DIGITAL UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; MONITOR,BP,WRIST,DIGITAL UNIT Back to Search Results
Catalog Number MDS4003
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "item 3704/medline deluxe digital wrist blood pressure monitor is defective because the battery has died 2x in the last 2 months.She also stated that the readings are incorrect, she stated that she received a reading elsewhere that was a different number than what the wrist monitor was reporting.".
 
Additional Manufacturer Narrative
The reporter stated that the wrist blood pressure monitor is defective because the battery has died 2x in the last 2 months and that the readings are incorrect; she received a reading elsewhere that was a different number than what the wrist monitor was reporting.Medline received the wrist bp monitor in used condition, inserted 2 aaa control batteries, and tested the sample according to the instructions for use using a fluke bp pump; two measurements were performed and results showed no indications of readings out of range.There was no indication of any malfunctions or power issues; the monitor turned on/off as normal and the cuff was properly inflated/deflated.The complaint was not confirmed, the cause code was updated to reflect unconfirmed defect - not confirmed with received sample, and the sample worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
MONITOR,BP,WRIST,DIGITAL UNIT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
shirisha mothe
three lakes drive
northfield, IL 60093--275
MDR Report Key50948709
Report Number1417592-2026-54650
Device Sequence Number18753629
Product Code DXN
UDI-Device Identifier00888277675493
UDI-Public00888277675493
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K122482
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS4003
Device Lot Number07523030001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/01/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2023
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-