• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Insulet Corporation Omnipod 5 Automated Insulin Delivery System; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Insulet Corporation Omnipod 5 Automated Insulin Delivery System; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number PT-001925
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 08/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the patient's blood glucose level rose to 18mmol/l (324 mg/dl) while wearing the pod between 25 and 36 hours.When the pod was removed from the infusion site (arm), the pod's cannula was found bent.As treatment, a new pod applied.
 
Additional Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed, and the product lot met all acceptance criteria.Locked down smartphone: lockdown.Omnipod software app version: 3.1.6.Operating system: n5004l-am-q-mv01602-06-01.06.Hardware: n5004l.Cgm sensor type: g7.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Omnipod 5 Automated Insulin Delivery System
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer (Section G)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
+1(978)600-7000
MDR Report Key50948725
Report Number3014585508-2026-72436
Device Sequence Number9552145
Product Code QFG
UDI-Device Identifier10385083000138
UDI-Public(01)10385083000138(11)260131(17)280131(10)PH1K01312631
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K203768
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPT-001925
Device Catalogue NumberPOD-OMNI-I1-6729
Device Lot NumberPH1K01312631
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/06/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured01/31/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-