| Model Number |
CF-HQ190L |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a therapeutic colonoscopy, the colonovideoscope's monitor image blacked out for 1-5 seconds.The procedure was completed using the same device.During the initial device evaluation, the scope's image was flickering.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer and identified during the inspection has been confirmed, as the device has evidence of fluid in the connector, causing a flickering image.The cause was determined to be related to a component failure.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|