Catalog Number MDS88200D
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that ¿the lift valve is closed and its not pumping up.¿ it was also reported that ¿the replacement hydraulic cylinder received in not exactly the same,¿ ¿this one does not seem to pump up high enough,¿ and ¿with the lever released, it does not drop the boom arm.¿.
 
Additional Manufacturer Narrative
The customer reported that the lift will not go up and the valve is closed; later, the customer stated that the replacement hydraulic cylinder received in not exactly the same, this one does not seem to pump up high enough, and with the lever released, it does not drop the boom arm; the complaint states the issue did not occur during use and no injury occurred.Medline reviewed photo evidence provided by the customer; a physical sample was not provided, and a functional or dimensional inspection cannot be executed via a photo.Investigation results were ¿confirmed bent boom, not confirmed leaking cylinder¿; after reviewing the photo, the boom arm appears to be severely bent downward, while the customer did not provide a photo or the sample of the original hydraulic cylinder.The root cause of this issue is unknown due to the lack of information provided by the customer; the investigation concluded it is likely that there was no issue with the hydraulic cylinder, but due to the damage to the boom, the customer was unable to raise and lower the lift to its usual height, and the division will monitor through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
LIFT,PATIENT,HYDRAULIC,6 POINT CRADLE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
joseph weiske
three lakes drive
northfield, IL 60093--275
MDR Report Key50948768
Report Number1417592-2026-54664
Device Sequence Number9624465
Product Code FSA
UDI-Device Identifier40080196764250
UDI-Public40080196764250
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS88200D
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2025
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-