The customer expressed concern about moisture between the inner and outer shell and suspected a possible crack in the inner shell causing leakage.Upon inspection, medline filled the unit with water and observed no leaks or visible cracks in the inner shell, and the pump was tested and is functioning properly.The complaint is not confirmed.No root cause was confirmed; the exact temperature of the water used by the customer is unknown, but it is possible that the introduction of ice water may have caused condensation between the inner and outer shell, and the division will continue to monitor this issue through trend analysis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|