| Model Number |
3570009 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a spectrum iq infusion pump failed downstream occlusion test during preventive maintenance.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.During functional testing, the customer reported ¿failed downstream occlusion test¿ was reproduced.The device was tested for flow testing for downstream occlusion testing and failed.A review of the event history log revealed downstream occlusion messages.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was verified.The cause of the condition was determined to be force sensor no tube calibration and force sensor scale factor calibration out of specification.The force sensor no tube and force sensor scale factor require calibration to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|