| Model Number |
BF-P190 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a bronchoscopy procedure, a little piece of the bottom by the lens of the bronchovideoscope came off and fell into the patient's lung.The broken piece was retrieved successfully with suction.No additional medical intervention or monitoring was required.There were no further reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.The device has evidence of c-cover came off and broken/big chip which is consistent with the reported event.Based on the results of the investigation, the definitive root cause of the reported issue could not be determined.The most probable cause of the complaint is cause traced to component failure, expected or random component failure without any design or manufacturing issue.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|