• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM? CRT-D; Cardiac resynchronization therapy defibrillator (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM? CRT-D; Cardiac resynchronization therapy defibrillator (CRT-D) Back to Search Results
Model Number G124
Medical Device Problem Codes High impedance (1291); Failure to Read Input Signal (1581); High Capture Threshold (3266)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited high pacing impedance measurements within range, high capture threshold, and low intrinsic amplitude.Technical services (ts) recommended further review.This device remains in system.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent changes in impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to description (b5) for more information regarding the specific circumstances of this event.Engineering analysis and testing of returned products has determined that repeated, small movements of the lead terminal ring along with variances in the connection contact force and surface roughness can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurement.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of pacing impedance high within normal limits was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM? CRT-D
Common Device Name
Cardiac resynchronization therapy defibrillator (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel
EI  
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50949002
Report Number2124215-2026-52044
Device Sequence Number15859062
Product Code NIK
UDI-Device Identifier00802526589102
UDI-Public(01)00802526589102(17)220120
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/20/2022
Device Model NumberG124
Device Catalogue NumberG124
Device Lot Number497630
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured01/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexMale
-
-