| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a monitor's response buttons were not functioning.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (response buttons non-functional) was confirmed.Upon investigation, the response button was not functional.The cause for the failure was caused by the response button dome being damaged.The root cause of the damaged dome cannot be positively identified.No adverse event resulted from the damaged monitor.
|
| |
|