Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 8mm force bipolar instrument was analyzed; however, the findings did not replicate or confirm the customer reported event.Manual inspection confirmed the instrument fully met all criteria, passed recognition and engagement tests on an in-house system, and demonstrated full range of motion in all directions.The grip tips opened and closed properly, and the instrument passed grip testing, energy delivery testing in various grip orientations, and electrical continuity testing.The instrument was fully functional.Additionally, the force bipolar instrument was found to have a broken grip tang at the middle of the grip, with an approximately 5mm x 2mm fragment confirmed to have detached from the instrument.The broken piece was not returned with the instrument.No further damage or abnormalities were identified.
|