The customer reported a hole punch out of glove and saw this in 3 different pairs.The complaint information was reviewed; inspection reports for the affected lot were reviewed, retained samples were checked for holes by visual and water leak test with no holes found, compounding batch records and dipping process control parameters were confirmed with no deviations identified, and the lot underwent 100% inspection for holes and visual defects using electronic pinhole testing with no abnormalities reported during pre-shipment release qa testing.After further review, a photo was received confirming the reported defect, and the vendor states that the defect was due to a machine process issue that was missed.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|