| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a monitor's response buttons were not functioning.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (response buttons non-functional) was confirmed.Upon investigation, the rear response button was not functional.As received, the rear response button trace was creased.The cause of the non-functional response buttons is the damaged trace.The root cause of the creased trace cannot be positively identified.No adverse event resulted from the damaged monitor.
|
| |
|