| Model Number |
SC-1216 |
| Medical Device Problem Code |
Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The spinal cord stimulation (scs) implantable pulse generator (ipg) database analysis performed identified a bluetooth fault code when charging the ipg.Noise from the charging field of the charger interferes with the bluetooth communication chipset, thus causing internal resets where the device stimulation momentarily turns off and then returns to normal operation.The ipg remains implanted in the patient and delivering therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
A field safety notice (fsn; z-2617-2024) was delivered to physicians in april 2024 communicating the potential for the wavewriter alpha 16 ipg spinal cord stimulation (scs) implantable pulse generator (ipg) stimulation therapy to be transiently suspended during charging due to a device reset.The device reset behavior occurs in response to the detection of potential noise/interference during ipg charging.When the system resets, the patients programmed stimulation is suspended for approximately 10-15 seconds and then the device returns to normal operation once the reset is complete, including the delivery of stimulation therapy.In december 2023, a firmware update was implemented that will prevent this device reset behavior from occurring during ipg charging.
|
| |
|