• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; STRIP,TEST,GLUCOSE,HARMONY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; STRIP,TEST,GLUCOSE,HARMONY Back to Search Results
Catalog Number MPH6550Z
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Incorrect, Inadequate or Imprecise Result or Readings (1535); Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "issues with blood glucose strips not reading correctly".
 
Additional Manufacturer Narrative
The customer reported issues with blood glucose strips not reading correctly; according to the facility the device "would sometimes lag and not get out of control mode" which would cause a delay in getting a reading and would at times "shut off after being turned on." medline received 2 samples, tested the meter using control blood glucose test strips and harmony glucose control solutions following the instructions for use, and completed four tests, two tests for each control solution.The complaint was not confirmed; during testing one of the meters turned on/off as normal, no warning or other error signs were displayed, and control solution readings were within range, while the other sample was visibly damaged as if it had been dropped and the middle of the screen would not display making it impossible to test and record readings.Investigation results were customer caused, the cause code was updated to reflect misuse/mishandling, and there are currently no known trends for this product lot and issue.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
STRIP,TEST,GLUCOSE,HARMONY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rylee rosenberg
three lakes drive
northfield, IL 60093--275
MDR Report Key50949548
Report Number1417592-2026-54744
Device Sequence Number9534366
Product Code NBW
UDI-Device Identifier20888277675688
UDI-Public20888277675688
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K161299
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH6550Z
Device Lot Number16823114001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/11/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2024
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-