Catalog Number DYNJ89561
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "we had 2 major surgery packs back-to-back that had pieces of black debris in them as well as one of them had a gown with some mold on it.".
 
Additional Manufacturer Narrative
The customer reported that they had 2 major surgery packs back-to-back that had pieces of black debris in them and one of them had a gown with some mold on it; the issue did not occur during use and no injury occurred.A photo and a sample were received for investigation, seven sample packs and a pack insert were provided, and the seven packs were individually opened with the interior components carefully removed; finished good production records and the finished good pack build were reviewed.Out of all seven packs, two were confirmed to have particles found within the 48oz bowl, one confirmed to be a corrugate particle and the other an unknown black particle; black and cardboard particles were found within the pack while the report of a dirty gown was not confirmed at this time.The original source of the black particles could not be identified, and the cardboard particle was small fragments of material originating from component packaging; the complaint was provided to the vendor for further evaluation and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-MAJOR SURGERY FMOL LADY OF THE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
lauren glaeser
three lakes drive
northfield, IL 60093--275
MDR Report Key50949686
Report Number1417592-2026-54757
Device Sequence Number10429259
Product Code FDE
UDI-Device Identifier10198459278297
UDI-Public10198459278297
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ89561
Device Lot Number25BBQ872
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/22/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2025
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-