The customer stated the back left brake came off the unit, it came right out of the handle after taking wife out of the chair, and the brake is now extremely loose and no longer engages; the complaint also states, "did an injury occur? :no." a physical sample was not provided; a photo depicting the defect was provided, and after reviewing the photo it was discovered that the left side brake cable appears to be disconnected from the hand brake assembly, the cable fracturing out of the handle.The investigation results were confirmed.Given the evidence provided in the photo, as well as the incident report, the root cause of this issue is unknown, and no trends were identified for the specified material number and its corresponding defect type.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|