Catalog Number MDS808210ARE
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "customer states back left brake came off the unit.Customer states it came right out of the handle after taking wife out of the chair.The brake is now extremely loose and no longer engages.".
 
Additional Manufacturer Narrative
The customer stated the back left brake came off the unit, it came right out of the handle after taking wife out of the chair, and the brake is now extremely loose and no longer engages; the complaint also states, "did an injury occur? :no." a physical sample was not provided; a photo depicting the defect was provided, and after reviewing the photo it was discovered that the left side brake cable appears to be disconnected from the hand brake assembly, the cable fracturing out of the handle.The investigation results were confirmed.Given the evidence provided in the photo, as well as the incident report, the root cause of this issue is unknown, and no trends were identified for the specified material number and its corresponding defect type.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TRANS CHAIR,ALUM,12 WHL, 19" 300 LB, RED
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
joseph weiske
three lakes drive
northfield, IL 60093--275
MDR Report Key50949722
Report Number1417592-2026-54761
Device Sequence Number16063870
Product Code INM
UDI-Device Identifier40884389150819
UDI-Public40884389150819
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS808210ARE
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-