• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Insulet Corporation Omnipod DASH Insulin Management System; Pump, Infusion, Insulin

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Insulet Corporation Omnipod DASH Insulin Management System; Pump, Infusion, Insulin Back to Search Results
Model Number 18320
Medical Device Problem Codes Loss of or Failure to Bond (1068); Unintended Movement (3026)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient's blood glucose level rose to greater than 250 mg/dl while wearing the pod between 5 and 24 hours.The adhesive completely separated from the (leg) causing the cannula to dislodge.As treatment, a new pod was inserted.
 
Additional Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed, and the product lot met all acceptance criteria.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Omnipod DASH Insulin Management System
Common Device Name
Pump, Infusion, Insulin
Manufacturer (Section D)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer (Section G)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
+1(978)600-7000
MDR Report Key50949735
Report Number3014585508-2026-72475
Device Sequence Number10446742
Product Code LZG
UDI-Device Identifier20385082000020
UDI-Public(01)20385082000020(11)260407(17)280407(10)PD1K04072621
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K211575
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number18320
Device Catalogue NumberBLE-I1-529
Device Lot NumberPD1K04072621
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured04/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-