| Catalog Number |
8975H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a aortic revision procedure on (b)(6)2026 and suture was used.During surgery (vascular procedure) the subject sutures were used, among other products.According to the surgeon, while suturing the anastomotic line, after several passes through the tissue, two of the sutures in use were torn in the middle section of the suture strand, and a third suture was torn in proximity to the needle attachment area, also after several passes through the tissue.It was further reported that a fourth suture may also have torn; however, the surgeon could not recall the details.The surgeon noted that the tissue was hard.All sutures that were torn originated from the same package.It should be noted that, according to the report, one or two additional sutures from the same package were used during the same procedure without a similar event occurring.The procedure was completed successfully, and there was no patient harm.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - pending evaluation of returned device.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.The following information was requested, but unavailable:
was there any reported patient or user harm?
was there a significant delay?
attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did the reported issue contribute to any patient adverse consequences?
do you have any photos available for visual analysis?
device history review: a manufacturing record evaluation was performed for the finished device batch tjbcux, (b)(6) and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|