• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Inc. PROLENE; suture, nonabsorbable, synthetic, polypropylene

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Inc. PROLENE; suture, nonabsorbable, synthetic, polypropylene Back to Search Results
Catalog Number 8975H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a aortic revision procedure on (b)(6)2026 and suture was used.During surgery (vascular procedure) the subject sutures were used, among other products.According to the surgeon, while suturing the anastomotic line, after several passes through the tissue, two of the sutures in use were torn in the middle section of the suture strand, and a third suture was torn in proximity to the needle attachment area, also after several passes through the tissue.It was further reported that a fourth suture may also have torn; however, the surgeon could not recall the details.The surgeon noted that the tissue was hard.All sutures that were torn originated from the same package.It should be noted that, according to the report, one or two additional sutures from the same package were used during the same procedure without a similar event occurring.The procedure was completed successfully, and there was no patient harm.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: g07002 - pending evaluation of returned device.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.The following information was requested, but unavailable: was there any reported patient or user harm? was there a significant delay? attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did the reported issue contribute to any patient adverse consequences? do you have any photos available for visual analysis? device history review: a manufacturing record evaluation was performed for the finished device batch tjbcux, (b)(6) and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE
Common Device Name
suture, nonabsorbable, synthetic, polypropylene
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.-San Lorenzo PR
road 183, km 8.3
san lorenzo PR 00754
PR   00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50949992
Report Number2210968-2026-09982
Device Sequence Number10446752
Product Code GAW
UDI-Device Identifier10705031022218
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8975H
Device Lot NumberTJBCUX
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/25/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-