• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; ELECTRODE, ECG, FOAM, 3PK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; ELECTRODE, ECG, FOAM, 3PK Back to Search Results
Catalog Number MDSM611503H
Medical Device Problem Codes Signal Artifact/Noise (1036); Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that updated packaged sku is less durable, button connectors are coming apart from pad and they are not sticking and generating a lot of artifact when used.
 
Additional Manufacturer Narrative
The customer reported the updated packaged sku is less durable, button connectors are coming apart from pad, they are not sticking, and generating a lot of artifact when used.Through internal investigation, including dhr review and inspection records review, no abnormalities were found; electrodes from retained lot were checked for fitment issues, and none of the electrodes demonstrated failures or irregularities.A sample was not received for investigation, and the complaint was not confirmed due to the fact that this issue can only be confirmed by further investigating the sample or a provided photo.The cause was sample required to determine cause, and based on the review of the details provided, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ELECTRODE, ECG, FOAM, 3PK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rylee rosenberg
three lakes drive
northfield, IL 60093--275
MDR Report Key50950260
Report Number1417592-2026-54802
Device Sequence Number9552190
Product Code DRX
UDI-Device Identifier10884389830321
UDI-Public10884389830321
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K041954
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDSM611503H
Device Lot Number10 65923110013
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2024
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-