It was reported that pump declared altitude alarm and insulin delivery was suspended, and customer¿s blood glucose level was displayed as ¿high¿ above 500 mg/dl.There was an adverse impact to the customer¿s blood glucose level, but specific value was not known.Customer initially took no action and did not seek third-party assistance, while cts provided instructions to clean speaker area, remove and reconnect cartridge, and attempt to resume insulin, which was not successful until customer loaded new cartridge and waited 2 hours for moisture to evaporate.After 2 hours, customer was able to load new cartridge, resume insulin delivery, and pump returned to normal operation, and cts explained likely exposure to condensation or humidity, gave tips to prevent altitude alarms, and closed case with no further action required.
|