Model Number 1000354
Medical Device Problem Code False Alarm (1013)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 08/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that pump repeatedly displayed altitude alarm and suspended insulin delivery over past month while within normal operating altitude.Customer¿s blood glucose level exceeded 500 mg/dl and meter displayed ¿high¿; specific value was unknown.Customer initially took no action for high blood glucose.Technical support guided customer to clean speaker holes with soft brush and lint-free cloth, remove and reconnect cartridge, and wait for insulin to resume, after which insulin delivery restarted and altitude alarm did not recur.Pump resumed normal operation, insulin remained active, and customer received instructions on preventing future altitude alarms.
 
Additional Manufacturer Narrative
In use at the time of the event: cgm model: unknown.Mobile os: unknown / unknown / unknown / unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
t:slim X2 Insulin Pump with Control-IQ+ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50950360
Report Number3013756811-2026-177969
Device Sequence Number9499355
Product Code QFG
UDI-Device Identifier00389152319810
UDI-Public(01)00389152319810(22)1012742(11)241217
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K232380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1000354
Device Catalogue Number1013198
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/17/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-