It was reported that pump repeatedly displayed altitude alarm and suspended insulin delivery over past month while within normal operating altitude.Customer¿s blood glucose level exceeded 500 mg/dl and meter displayed ¿high¿; specific value was unknown.Customer initially took no action for high blood glucose.Technical support guided customer to clean speaker holes with soft brush and lint-free cloth, remove and reconnect cartridge, and wait for insulin to resume, after which insulin delivery restarted and altitude alarm did not recur.Pump resumed normal operation, insulin remained active, and customer received instructions on preventing future altitude alarms.
|