The account reported contrast controller leak, and the complaint described leaking air out of the top of them, difficult time filling the buritrol, and a hissing sound.Eight new, unused samples were returned for evaluation; visual evaluation did not reveal any defects, and functional evaluation was performed by connecting the spike to a saline bag and squeezing, with results found the samples to function as intended.Based on the investigation results, the reported event was not able to be confirmed at this time.Root cause remains inconclusive at this time as the reported defect was not able to be confirmed, and no action is required at this time as the reported event was not able to be confirmed.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|