It was reported that pump declared cartridge change error without alarm, and customer¿s blood glucose level exceeded 500 mg/dl with value displayed as ¿high.¿ customer experienced adverse impact to blood glucose level due to reported issue.Customer gave correction injection using multiple daily injections and did not require help from another person, then, with guidance from technical support, inspected cartridge o-ring and pump area, cleaned pneumatic tap, reattached cartridge, and followed on-screen steps to complete load process, after which customer successfully resumed insulin delivery.
|