| Model Number |
CF-HQ190L |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a fuzzy image with white specks during the pre-use inspection of an olympus colonovideoscope.During the initial device evaluation, a noisy image condition was confirmed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The reported failure was confirmed.A definitive root cause could not be established.However, the investigation confirmed a noisy image condition and identified a defective plug unit.Therefore, the most probable cause of the reported fuzzy image with white specks was damage to the plug unit, resulting in degraded image quality.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|