The reported event of inappropriate or unexpected reset was confirmed.The device voltage was above elective replacement indicator (eri) upon receipt.Assessment of total projected longevity was performed, and the device was found to have appropriate remaining longevity.Analysis of the device image indicated reset error occurred, consistent with time of event device having been dropped in the field; however, the error condition is no longer present.User manual states sterility, integrity, or function cannot be guaranteed upon device being dropped.Further analysis of the device¿s telemetry, lead impedance, pacing, sensing, and high voltage charging were found to be normal.The backup operation could not be reproduced during analysis.
|