| Model Number |
GIF-HQ190 |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the gastrointestinal videoscope exhibited something within the air/water channel, with some kind of debris observed at the distal tip during inspection for use.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, olympus confirmed white residue came out of the nozzle, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|