| Model Number |
3660 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the patient system was unable to disable mri mode following an mri.An abbott representative met with the patient and was able to successfully disable the patient¿s ipg from mri mode using an orion exit tool (oet).Once the ipg was taken out of mri mode, the patient controller and/or clinician programmer were able to bond with the ipg, and restored therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated
based on information obtained the device is included in the neuromodulation implantable pulse generator (ipg) unable to exit mri mode advisory notice issued by abbott on (b)(6)2023.Based on the information received, the event is consistent with the loss of the bluetooth pairing bond while in mri mode.As a result of this finding, abbott is performing further investigation.
|
| |
|