| Model Number |
SB-0535FC |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the instrument to the service center for a separate issue.During the device analysis, the device evaluator indicated that the probe was broken off.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, reasonably known information suggests probable causes such as damage or deterioration of parts due to wear and tear, without any design or manufacturing issue.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.A review of the device history record identified no process deviations or nonconformities that could have contributed to the reported event.Also, it was confirmed that the device was shipped in compliance with specified requirements.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|