• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; ANEROID,HAND HELD,ADULT,NEOPRENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; ANEROID,HAND HELD,ADULT,NEOPRENE Back to Search Results
Catalog Number MDS9380LF
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "we have a defective adult bp cuff that does not inflate properly.".
 
Additional Manufacturer Narrative
The reported issue was a defective adult bp cuff that does not inflate properly.Medline received 1ea medline handheld aneroid in used condition and proceeded to test the sample according to the instructions for use; the sample was inflated up to 300 mmhg, and while inflating/deflating the sample, the gauge displayed measurements without any issues.The test was performed numerous times and no issues were experienced; the complaint was not confirmed.The sample worked as intended, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ANEROID,HAND HELD,ADULT,NEOPRENE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
joachim kamanda
three lakes drive
northfield, IL 60093--275
MDR Report Key50951875
Report Number1417592-2026-55043
Device Sequence Number15836299
Product Code DXQ
UDI-Device Identifier10080196738922
UDI-Public10080196738922
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K211084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS9380LF
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/11/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-