• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Insulet Corporation Omnipod DASH Insulin Management System; Pump, Infusion, Insulin

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Insulet Corporation Omnipod DASH Insulin Management System; Pump, Infusion, Insulin Back to Search Results
Model Number 18320
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Omnipod DASH Insulin Management System
Common Device Name
Pump, Infusion, Insulin
Manufacturer (Section D)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer (Section G)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
+1(978)600-7000
MDR Report Key50952146
Report Number3014585508-2026-72574
Device Sequence Number13145225
Product Code LZG
UDI-Device Identifier20385082000020
UDI-Public(01)20385082000020(11)251009(17)271009(10)PD1K10092511
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K211575
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number18320
Device Catalogue NumberBLE-I1-529
Device Lot NumberPD1K10092511
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/28/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-