date sent: 10/06/2026.Additional information was requested, and the following was obtained:
1.Please provide device details lot#/batch# don¿t have the device at this point
2.Did the device lock-out (no staples deployed and no cut line started)? no
3.Or, did the device start to deploy staples and cut but could not be completed? yes
4.Or, did the device fully fire and stop during the automatic knife return? no
5.Was there any difficulty opening the device jaws? manual override was engaged.The knife blade was fully returned.The stapler was opened
6.If yes, what steps were taken to open the jaws.Manual override was engaged.The knife blade was fully returned.The stapler was opened
7.If the jaws could not be opened, how was the device removed.Na