It was reported that guidewire stuck, guidewire kink, guidewire lumen obstruction and deployment issue occurred.During an atrial fibrillation ablation procedure, a farawave catheter was selected for use.Multiple attempts to remove the guidewire from the catheter were unsuccessful.Troubleshooting efforts, including advancing and backloading the guidewire, were performed; however, resistance was encountered and the guidewire lumen appeared to be obstructed.Farawave catheter deployment issue was also observed during use.Farawave catheter was subsequently removed from the patient.Upon removal, the guidewire was noted to be kinked.Farawave catheter was replaced with a new farawave catheter to complete the procedure without further issue.No patient complications were reported.It was further reported the splines were reported to be out of plane when the deployment issue was noted.
|