The reported issue was low suction/inadequate pressure/output/suction.Medline quality received one unit, visually inspected the sample, plugged the sample into a power source and turned the unit on, adjusted the pressure with the pressure dial, measured 600 mmhg at maximum, and opened the sample to check for internal damages and defects and found none.The reported issue was not confirmed with the received sample, and no root cause was identified.The unit appeared to be in the original packaging and unused, appeared to be working and within product specifications, and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|