MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) ASSURITY MRI; Implantable Pulse Generator, Pacemaker (Non-Crt)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) ASSURITY MRI; Implantable Pulse Generator, Pacemaker (Non-Crt) Back to Search Results
Model Number PM2272
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient exhibited twiddler¿s syndrome causing migration of the pacemaker and dislodgement of the right atrial (ra) and right ventricular (rv) leads.The dislodgement resulted in diaphragmatic stimulation and patient discomfort.Programming changes were made to address the issue.On (b)(6) 2026, during the procedure, the physician explanted the right atrial (ra) lead.A replacement ra lead was installed; however, the lead became dislodged and exhibited noise, high impedance, and low current of injury.Ra lead fracture was suspected.This ra lead was not used, and the physician completed the procedure using a different ra lead replacement.Additionally, the rv lead was explanted and replaced during the procedure.The patient was discharged, same day, following the procedure.
 
Additional Manufacturer Narrative
Further information was requested but not received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Common Device Name
Implantable Pulse Generator, Pacemaker (Non-Crt)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50952390
Report Number2017865-2026-20765
Device Sequence Number18547731
Product Code LWP
UDI-Device Identifier05414734509589
UDI-Public(01)05414734509589(10)A000182972(17)270630
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P140033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPM2272
Device Lot NumberA000182972
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/26/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-