• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) UltiPace¿ Lead, 46cm MRI; Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) UltiPace¿ Lead, 46cm MRI; Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes Back to Search Results
Model Number LPA1231/46
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Use of Device Problem (1670); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient exhibited twiddler¿s syndrome causing migration of the pacemaker and dislodgement of the right atrial (ra) and right ventricular (rv) leads.The dislodgement resulted in diaphragmatic stimulation and patient discomfort.Programming changes were made to address the issue.On (b)(6) 2026, during the procedure, the physician explanted the right atrial (ra) lead.A replacement ra lead was installed; however, the lead became dislodged and exhibited noise, high impedance, and low current of injury.Ra lead fracture was suspected.This ra lead was not used, and the physician completed the procedure using a different ra lead replacement.Additionally, the rv lead was explanted and replaced during the procedure.The patient was discharged, same day, following the procedure.
 
Additional Manufacturer Narrative
Further information was requested but not received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UltiPace¿ Lead, 46cm MRI
Common Device Name
Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50952393
Report Number2017865-2026-20768
Device Sequence Number9534047
Product Code NVN
UDI-Device Identifier05415067046345
UDI-Public(01)05415067046345(10)P000260011(17)290630
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P960013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLPA1231/46
Device Catalogue NumberLPA1231/46
Device Lot NumberP000260011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/09/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-